• Font Size
  • S
  • M
  • L
  • No : 4743
  • Date of Issue : 2021/01/08 13:58
  • Print

LC pump cannot solvent delivery (LCMS)

LC pump cannot solvent delivery (LCMS)
Category : 

Answer

It is set to turn on cooperation Nebulizing Gas and Pump.
The [Gas Control Mode] tab page is located in the [Setting Options] sub-window. To open this sub-window, click [Option] on the [Tools] menu in the [MS Data Analysis] window.
If this item is selected, the LC pump is shut OFF automatically when the nebulizer gas stops with a mass spectrometer specified as connected in system configuration settings.
 
NOTE: When the LC pump stops, the Nebulizing Gas is not stopped.
With Micro-ESI or PESI inteface, this parameter is disabled.

May I have your opinion?

Please provide your feedback below. This is for feedback only. You will not receive a response.
This site uses the FAQ system "PKSHA FAQ", which is a product of PKSHA Communication Inc.