FAQ - Frequently Asked Questions
- Products Search > GC-MS > Troubleshooting > (GCMS) "Detector saturated" message appears during analysis
(GCMS) "Detector saturated" message appears during analysis
(GCMS) "Detector saturated" message appears during analysis.
- Category :
Answer
1. Turn on the filament after solvent has completely eluted. If peaks elute before the solvent, use a time program to turn the filament off and on.
2. Decrease the detector voltage.
3. Reduce the sample concentration by making a dilution or reduce the injection volume.
4. If unavoidable, switch the filament OFF.
May I have your opinion?
This site uses the FAQ system "PKSHA FAQ", which is a product of PKSHA Technology Inc.