(GCMS) Autotuning and Analysis Errors

(GCMS) Autotuning and Analysis Errors
Category : 

Answer

1. Error message displayed when standard is introduced.
  Checks that Filament ON Errors.
 
2. "Initialization Error" message appears.
  When standard not introduced, open PFTBA standard valve.
  Click here for Standard Reagent (PFTBA) Addition Procedure.
 
3. "Resolution Adjustment Error", "Intensity Adjustment Error", "Mass Calibration Error", "Mass Pattern Correction Error" message appears.
  An unknown error occurred during tuning. Perform the tuning again.
  In the same error appears when tuning is repeated, clean the ion source and repeller.
  Click here for EI Ion Source Replacement and Rinsing Procedure.
  When the same error still appears after taking the measures above, contact a Shimadzu sales or service representative.