FAQ - Frequently Asked Questions
- Products Search > GC-MS > Troubleshooting > (GCMS) Chemical Ionization(CI, NCI) When good spectrum ca...
(GCMS) Chemical Ionization(CI, NCI) When good spectrum cannot be obtained
(GCMS) Chemical Ionization(CI, NCI) When good spectrum cannot be obtained
- Category :
Answer
There may be leak of the reagent gas if the conditions are not satisfied in spite that the reagent gas is being supplied. The possible causes of leak are described below:
Cause 1: Either the cylinder valve of the gas cylinder, the control knob of the pressure regulator, or the outlet valve of the pressure regulator is closed. The solenoid valve of the reagent gas control unit is set to OFF.
Action 1: Check each open/close condition. Open it or set it to ON if there is any closed or OFF valve.
Cause 2: The ion source box is not mounted correctly or the cap of the interface is not inserted completely into the ion source box.
Action 2: Stop the instrument and check the state of the ion source. Be sure to turn OFF the power of the instrument.
Action 2: Stop the instrument and check the state of the ion source. Be sure to turn OFF the power of the instrument.
Cause 3: The PTFE tube of the ion source is not connected or poorly attached.
Action 3: Stop the instrument and check how the PTFE tube is attached to the ion source.
Be sure to turn OFF the power supply.
Action 3: Stop the instrument and check how the PTFE tube is attached to the ion source.
Be sure to turn OFF the power supply.
Click here for CI, NCI Ion Source Replacement and Rinsing Procedure.
Cause 4: Correct type of ion source box is not mounted. Correct repeller electrode spacer is not mounted.
Action 4: Stop the instrument and check the type of ion source and the type of repeller electrode spacer. Be sure to turn OFF the power of the instrument.
Action 4: Stop the instrument and check the type of ion source and the type of repeller electrode spacer. Be sure to turn OFF the power of the instrument.
Click here for CI, NCI Ion Source Replacement and Rinsing Procedure.
Cause 5: When DI is in use, the tip of the sample inlet probe is not inserted inlet hole of the ion source or incompletely inserted.
Action 5: Check if the sample inlet probe is at the insertion position. Perform position adjustment of the sample inlet.
Click here for DI-2010 Operation Methods.
If none of the above causes applies, contact a Shimadzu sales or service representative
May I have your opinion?
This site uses the FAQ system "PKSHA FAQ", which is a product of PKSHA Communication Inc.