The temperature inside the cabinet is at least 10degree-C higher than room temperature

The temperature inside the cabinet is at least 10ºC higher than room temperature.
Category : 

Answer

Did the temperature control function stop?
 
Models with temperature control functionality are designed with higher insulation performance inside the cabinet. If the temperature control functionality is not used, heat generated from the needle drive motor or other components can increase the temperature in the cabinet to 10ºC or more above room temperature.
Always use the temperature control mode on models that include temperature control functionality.
 