The rinse solution volume in the seal automatic rinsing kit increases

The rinse solution volume in the seal automatic rinsing kit increases.
Category : 

Answer

Is there leakage from the plunger seal?
Replace the plunger seal. If the plunger seal service life was extremely short, then also replace the plunger.