  • No : 4152
  • Date of Issue : 2020/12/02 16:46
  • Print

Error Code:[0x5292] ERROR FCVx DISCONNECED

Error Message: Autosampler: Communication with FCV disconnected during analysis.
Category : 

Answer

«Main Cause»
The cause is that communication with the FCV whose [FCV FIX] setting is [1: ON] is cut off.
 
«Countermeasure»
Make sure the FCV is turned on. Check if the option cable is disconnected.
If it is not necessary to detect this error, set [FCV FIX] to [0: OFF].