FAQ - Frequently Asked Questions
- Products Search > HPLC > Autosampler > Error Code:[0x5291] ERROR ACQUISITION TIME
Error Code:[0x5291] ERROR ACQUISITION TIME
Error Message: Autosampler: Repeated injections were not completed within the data acquisition time.
- Category :
Answer
«Main Cause»
Autosampler: Repeated injections did not complete within the data acquisition time.
«Countermeasure»
Set the injection interval to a time longer than the actual injection operation time, and change the LC stop time and detector's data acquisition time to a time longer than "Injection interval × number of injections".
May I have your opinion?
This site uses the FAQ system "PKSHA FAQ", which is a product of PKSHA Technology Inc.