• Font Size
  • S
  • M
  • L
  • No : 4151
  • Date of Issue : 2020/12/02 16:46
  • Print

Error Code:[0x5291] ERROR ACQUISITION TIME

Error Message: Autosampler: Repeated injections were not completed within the data acquisition time.
Category : 

Answer

«Main Cause»
Autosampler: Repeated injections did not complete within the data acquisition time.
 
«Countermeasure»
Set the injection interval to a time longer than the actual injection operation time, and change the LC stop time and detector's data acquisition time to a time longer than "Injection interval × number of injections".

May I have your opinion?

Please provide your feedback below. This is for feedback only. You will not receive a response.
This site uses the FAQ system "PKSHA FAQ", which is a product of PKSHA Communication Inc.