Error Code:[0x5291] ERROR ACQUISITION TIME

Error Message: Autosampler: Repeated injections were not completed within the data acquisition time.
Category : 

Answer

«Main Cause»
Autosampler: Repeated injections did not complete within the data acquisition time.
 
«Countermeasure»
Set the injection interval to a time longer than the actual injection operation time, and change the LC stop time and detector's data acquisition time to a time longer than "Injection interval × number of injections".