Error Code:[0x528B] WARNING SMP, L: -

Error Message: Autosampler: Sample table is not correct.
Category : 

Answer

«Main Cause»
An error occurred in the sample table being executed. The row number of the row with an error is indicated to the right of "L:" and the error type is indicated to the right of the hyphen.
 
«Countermeasure»
Fix the error indicated for the row number in the message.
PLATE: Plate No.
FromSN: Starting vial number
ToSN:Ending vial number