Error Code:[0x5256] ERROR PFILE WAIT.RT RUN

Error Message: Autosampler: The WAIT.RT command of the pretreatment program is set prior to start analysis. Correct the pretreatment program.
Category : 

Answer

«Main Cause»
An analysis time was specified before the analysis is started by the pretreatment program.
 
«Countermeasure»
Revise the pretreatment program so that the WAIT.R command is executed after the analysis starts.