• Font Size
  • S
  • M
  • L
  • No : 3946
  • Date of Issue : 2020/12/02 16:44
  • Print

Error Code:[0x5241] ERROR PFILE MIX POS

Error Message: Autosampler: The MIX command of the pretreat program is not able to executed. Correct the pretreatment program.
Category : 

Answer

«Main Cause»
The needle is not positioned inside or above a sample vial when the MIX command is executed in the pretreatment program.
 
«Countermeasure»
Add a VIAL.N or other command prior to the MIX command to move the needle to the sample vial to be mixed.

May I have your opinion?

Please provide your feedback below. This is for feedback only. You will not receive a response.
This site uses the FAQ system "PKSHA FAQ", which is a product of PKSHA Communication Inc.