• Font Size
  • S
  • M
  • L
  • No : 3918
  • Date of Issue : 2020/12/02 16:44
  • Print

Error Code:[0x5225] ERROR NO PLATE

Error Message: Autosampler: The Sample plate is not set to the position set in the sequence. Set sample plate to the correct position.
Category : 

Answer

«Main Cause»
No sample plate is located at the tray number specified in the sequence.
 
«Countermeasure»
Place a sample plate in the tray number position specified in the sequence.
If [VIAL SKIP] is set to [Enable] in system settings, then analysis is skipped for positions without a sample vial and analysis continues from the next vial.

May I have your opinion?

Please provide your feedback below. This is for feedback only. You will not receive a response.
This site uses the FAQ system "PKSHA FAQ", which is a product of PKSHA Communication Inc.