Error Code:[0x5224] ERROR NO VIAL

Error Message: Autosampler: The vial is not set to the position set in the sequence. Set sample vial to the correct position.
Category : 

Answer

«Main Cause»
No sample vial is placed at the vial number position specified in the sequence.
 
«Countermeasure»
Place a sample vial at the vial number position specified in the sequence.
If [VIAL SKIP] is set to [Enable] in system settings, then analysis is skipped for positions without a sample vial and analysis continues from the next vial.