Error Code:[0x521A] FATAL INNER FAN

Error Message: Autosampler: The temperature control was stopped due to the detection of a rotation error for a fan in the cabinet. Check the fan for obstacles.
Category : 

Answer

«Main Cause»
A fan rotation problem was detected for the fan inside the cabinet.
 
«Countermeasure»
Check for foreign matter caught in the fan in the cabinet. If that does not resolve the error, switch the power OFF and contact a Shimadzu sales or service representative.