- Products Search > HPLC > Solvent Delivery Module > Error Code:[0x81E8] ERROR NO CONST.PRS
Error Code:[0x81E8] ERROR NO CONST.PRS
Error Message: Pump: The pumping was stopped because of settings of unsupported constant-pressure pumping mode. Use the constant flow pumping mode.
- Category :
Answer
«Main Cause»
The system controller specified the constant-pressure pump mode while a solvent delivery pump without constant-pressure pumping capability is being used.
«Countermeasure»
Use the constant-flowrate pumping mode.