Error Code:[0x5D14]

Error Message: Mobile phase monitor: The zero-point value is far out of the reference value. Perform zero-point registration without placing anything on the bottle holder.
Category : 

Answer

«Main Cause»
The zero point value has deviated significantly from the reference value.
 
«Countermeasure»
Please ensure there are no objects placed on the bottle holder and perform zero-point registration again. If the error still occurs, unplug the power cord for the main unit and contact a Shimadzu sales or service representative.