FAQ - Frequently Asked Questions
Error Code:[0x8720] ERR FAN STOP
Error Message: Cooling fan for the Xe lamp power has stalled.
- Category :
Answer
«Main Cause»
Displayed when the cooling fan for the Xe lamp's power supply stops.
«Countermeasure»
Check if the cooling fan is working . If the exhaust vents at the rear and side are obstructed, secure sufficient space for them to operate.
If the filters in the front cover and right side face are clogged, replace these filters. If the same error is still displayed, contact a Shimadzu sales or service representative.
When this error occurs, the temperature adjustment function of flow cell is forcibly turned. off. To turn the function on again, set the temperature for [CELL TEMP] in the parameter settings group.
May I have your opinion?
This site uses the FAQ system "PKSHA FAQ", which is a product of PKSHA Communication Inc.