Error Code:[0x871F] ERR PELTIER FAN

Error Message: Fan for the Peltier unit has stalled.
Category : 

Answer

«Main Cause»
Displayed when the cooling fan for the Peltier cooling fan has stopped.
 
«Countermeasure»
Turn the power OFF and then restart the instrument. Set the temperature again for [CELL TEMP] in the parameter settings group. If the same error is still displayed, contact a Shimadzu sales or service representative.