Error Code:[0x871D] ERR CELL TEMP
Error Message: Flow cell is overheated.
- Category :
Answer
«Main Cause»
Displayed when the flow cell temperature has risen abnormally.
«Countermeasure»
If the exhaust vents at the rear and side are obstructed, secure sufficient space for them to operate. If the filters in the front cover and right side face are clogged, replace these filters. If the error is still displayed, turn the power OFF and contact a Shimadzu sales or service representative.
When this error occurs, the temperature adjustment function of the flow cell is forcibly turned off. To turn the function on again, set the temperature for [CELL TEMP] in the parameter setting group.