- Products Search > HPLC > Supercritical Fluid Chromatograph > Error Code:[0x8B07] SFE ERR DISCONNECT
Error Code:[0x8B07] SFE ERR DISCONNECT
Error Message: Rack changer connect disconnection error
- Category :
Answer
«Main Cause»
The connection cable between the SFE and the rack-change is not connected properly. Or the rack-changer is turned off.
«Countermeasure»
Check the rack-changer is turned on.
If the rack-changer is turned on, check the RS-232C cable between the rack-changer and the SFE is connected properly.