Error Code:[0x89C3]

Error Message: D2 Lamp current is too low.
Category : 

Answer

«Main Cause»
Probable causes are the D2 lamp connector is not connected, the D2 lamp malfunctioned, or the lighting circuit malfunctioned.
 
«Countermeasure»
Check the D2 lamp connections. Check the total operating time of the lamp, and replace the lamp.
Click here for Replacement method of Lamp (HPLC)
If this does not remedy the problem, contact a Shimadzu sales or service representative.