Error Code:[0x8965]
Error Message: Cell is not calibrated. Register cell and adjustment and wavelength calibration the exposure time.
- Category :
Answer
«Main Cause»
(1) An unregistered cell was installed in the instrument.
(2) Failed to read cell ID.
«Countermeasure»
(1) Perform cell registration, exposure time adjustment, and wavelength calibration.
(2) Check the connection of the cell cable.