- Products Search > HPLC > Solvent Delivery Module > Error Code:[0x0030] ERROR P.MAX
Error Code:[0x0030] ERROR P.MAX
Error Message: Solvent delivery unit :pump's pressure has exceeded P.Max. Check for flow line clogging and P.Max Settings.
- Category :
Answer
«Main Cause»
The discharge pressure of pump exceeds the Settings [P.Max]. Pumping will automatically stop. However, if "S-PROT" of auxiliary operation is Settings, set flow rate to 1/2 for pumping.
«Countermeasure»
Inspection the flow line clog. If there is no \n abnormality, Settings the appropriate values in [P.Max].