• Font Size
  • S
  • M
  • L
  • No : 3634
  • Date of Issue : 2020/12/02 16:30
  • Print

Error Code:[0x0031] ERROR P.MIN

Error Message: The pressure of solvent delivery unit :pump does not reach P.Min. Check flow line leak and P.Min Settings.
Category : 

Answer

«Main Cause»
Pump discharge pressure has fallen below the set min. pressure limit, [P.MIN]. Pumping will automatically stop.
 
«Countermeasure»
Flow line fluid leakage or aspiration of air bubble is cause. Inspection between flow line and mobile phase. If there is no abnormality, Settings the appropriate values in [P.MIN]. Note that [P.MIN] function is not activated for first one minute of pumping.

May I have your opinion?

Please provide your feedback below. This is for feedback only. You will not receive a response.
This site uses the FAQ system "PKSHA FAQ", which is a product of PKSHA Communication Inc.