Error Code:[0x81D4] OPEN DRAIN VALVE
Error Message: The pressure is very high when you are purge.
- Category :
Answer
«Main Cause»
Pump discharge pressure exceeded [P-PMAX] when purge.
«Countermeasure»
Open drain valve and press [Purge] button. If an error message appears during auto-purging by system controller, inspect for clogging in the flow line. If no clogging is found, set the appropriate [P-FLOW] and [P-PMAX] value.