Error Code:[0x85F1] REMOVE RACK

Error Message: Rack is double-set. Remove the rack or turn it dummy rack. The schedule will end after the analysis of the rack. Please precaution it.
Category : 

Answer

«Main Cause»
The same rack has been set twice. After analysis is performed for this rack, [RC FULL RACK] error is displayed and the scheduled operation is stopped.
 
«Countermeasure»
Either removes the rack from the rack position for which the indicator is flashing, or replaces it with a condensation-prevention rack.