  • No : 3623
  • Date of Issue : 2020/12/02 16:30
  • Print

Error Code:[0x85EF] SET RACK

Error Message: Some stacks do not have rack set. Please precaution your schedules when you analysis the rack.
Category : 

Answer

«Main Cause»
There is a stack for which a rack has not been set. When analysis is performed for this rack, [RC NO. RACK] error is displayed and the scheduled operation is stopped.
 
«Countermeasure»
Set a rack in the rack position in the queue for analysis. When the autosampler is quipped with cooling function, set racks with empty microtiter plates or deep-well plates in all unused stacks to prevent dew condensation.