- Products Search > HPLC > Autosampler > Error Code:[0x85EF] SET RACK
Error Code:[0x85EF] SET RACK
Error Message: Some stacks do not have rack set. Please precaution your schedules when you analysis the rack.
- Category :
Answer
«Main Cause»
There is a stack for which a rack has not been set. When analysis is performed for this rack, [RC NO. RACK] error is displayed and the scheduled operation is stopped.
«Countermeasure»
Set a rack in the rack position in the queue for analysis. When the autosampler is quipped with cooling function, set racks with empty microtiter plates or deep-well plates in all unused stacks to prevent dew condensation.