- Products Search > HPLC > Autosampler > Error Code:[0x85E8] RC FULL RACK
Error Code:[0x85E8] RC FULL RACK
Error Message: You cannot set rack to rack position the during analysis of rack changer.
- Category :
Answer
«Main Cause»
While a rack is being analyzed, another rack is set in its rack position.
«Countermeasure»
Remove the illegal rack or change it to the condensation-prevention rack in the analysis rack position whose indicator is flashing, and turn OFF and ON the power to the instrument.