- Products Search > HPLC > Autosampler > Error Code:[0x85E0] RC NO RACK
Error Code:[0x85E0] RC NO RACK
Error Message: The schedule analysis was ended because the rack was not set.
- Category :
Answer
«Main Cause»
There was no rack in the position specified for analysis and so the scheduled analysis was stopped.
«Countermeasure»
Set a rack in a position whose indicator is flashing, and turn OFF and ON the power to the instrument.
If the error occurs persistently, the rack position or condition has changed before and after the power is turned ON. Execute [Clear RACK INFO] on the autosampler to delete the rack information, and turn OFF and ON the power to the instrument.