Error Code:[0x8548] STOP COOLER R

Error Message: The rack has not been set for more than 30 minutes in the rack plate R, so the rack plate R cooler was turned off. Set rack to restart the sample cooler cell temperature control.
Category : 

Answer

«Main Cause»
Since sample cooler R has operated without sample rack for more than 30 minutes, sample cooler R is stopped.
 
«Countermeasure»
Insert a sample rack into rack plate R. Or, stop sample cooler R.

May I have your opinion?