• Font Size
  • S
  • M
  • L
  • No : 3597
  • Date of Issue : 2020/12/02 16:30
  • Print

Error Code:[0x8547] STOP COOLER M

Error Message: The rack has not been set for more than 30 minutes in the rack plate M, so the rack plate M cooler was turned off. Set rack to restart the sample cooler cell temperature control.
Category : 

Answer

«Main Cause»
Since sample cooler M has operated without sample rack for more than 30 minutes, sample cooler M is stopped.
 
«Countermeasure»
Insert a sample rack into rack plate M. Or, stop sample cooler M.

May I have your opinion?

Please provide your feedback below. This is for feedback only. You will not receive a response.
This site uses the FAQ system "PKSHA FAQ", which is a product of PKSHA Technology Inc.