Error Code:[0x001E] SIL:SYRINGE VOLUME
Error Message: The Settings of aspiration /discharge in the measuring pump of pretreatment program is out of range.
- Category :
Answer
«Main Cause»
Invalid aspiration/discharge volume setting of metering pump in pretreatment program.
«Countermeasure»
If suction volume is larger than syringe capacity, reset it to a smaller volume than syringe capacity.
If discharge volume is larger than suction volume, reset discharge volume to a smaller volume than sucked volume, or reset suction volume to a larger