FAQ - Frequently Asked Questions
- Products Search > HPLC > Autosampler > Error Code:[0x001C] SIL:SYRINGE VOLUME
Error Code:[0x001C] SIL:SYRINGE VOLUME
Error Message: The Settings of aspiration /discharge of pretreatment program is out of range.
- Category :
Answer
«Main Cause»
Suction of more volume than syringe capacity was attempted in pretreatment program.
«Countermeasure»
If suction volume is larger than syringe capacity, reset it to a smaller volume than syringe capacity.
If discharge volume is larger than suction volume, reset discharge volume to a smaller volume than sucked volume, or reset suction volume to a larger
May I have your opinion?
This site uses the FAQ system "PKSHA FAQ", which is a product of PKSHA Communication Inc.