• Font Size
  • S
  • M
  • L
  • No : 3468
  • Date of Issue : 2020/12/02 16:10
  • Print

Error Code:[0x9202]

Error Message: The leak sensor of the autosampler high pressure valve has detected a liquid leak.
Category : 

Answer

«Main Cause»
There is a liquid leak from the high pressure valve, or from a unit positioned above the high pressure value (column oven or detector, etc.).
 
«Countermeasure»
Inspect the liquid leak from the high pressure valve, or from a unit positioned above the high pressure value (column oven or detector, etc.), and wipe away any liquid from around the leak sensor.

May I have your opinion?

Please provide your feedback below. This is for feedback only. You will not receive a response.
This site uses the FAQ system "PKSHA FAQ", which is a product of PKSHA Communication Inc.