Error Code:[0x9201]

Error Message: The leak sensor of the autosampler low pressure valve has detected a liquid leak.
Category : 

Answer

«Main Cause»
There is a liquid leak from the low pressure valve or metering pump, or from a unit positioned above the low pressure value (high pressure valve, column oven or detector, etc.).
 
«Countermeasure»
Inspect the liquid leak from the low pressure valve or metering pump, or in a unit positioned above the low pressure value (high pressure valve, column oven or detector, etc.), and wipe away any liquid from around the leak sensor.