Error Code:[0x93CC]
Error Message: The detector W lamp is not connected.
- Category :
Answer
«Main Cause»
The W lamp connector is not connected.
Alternatively, the W lamp cable is broken or there is a fault in the board.
Alternatively, the W lamp cable is broken or there is a fault in the board.
«Countermeasure»
Follow the procedures below in the following order.
(1) Connect the W lamp connector.
(2) Replace the W lamp.
(1) Connect the W lamp connector.
(2) Replace the W lamp.
Click here for Replacement method of Lamp (HPLC)
(3) If the error is not resolved, press the (power switch) to turn the power OFF and contact a Shimadzu sales or service representative.