Error Code:[0x93CB]
Error Message: The detector D2 lamp is not connected.
- Category :
Answer
«Main Cause»
The D2 lamp connector is not connected.
Alternatively, the D2 lamp cable is broken or there is a fault in the board.
Alternatively, the D2 lamp cable is broken or there is a fault in the board.
«Countermeasure»
Follow the procedures below in the following order.
(1) Check the connection with the D2 lamp.
(2) Replace the D2 lamp.
(1) Check the connection with the D2 lamp.
(2) Replace the D2 lamp.
Click here for Replacement method of Lamp (HPLC)
(3) Press the (power switch) to turn the power OFF and contact a Shimadzu sales or service representative.