Error Code:[0x93C6]

Error Message: The detector lamp housing cover is open. Turn the lamp off.
Category : 

Answer

«Main Cause»
The lamp house cover is open or there is a fault in the cover sensor.
 
«Countermeasure»
Close the lamp house cover.
If the error is not resolved, replace the cover sensor.