Error Code:[0x93C3]
Error Message: The D2 lamp cannot light because the lamp current of the detector is too small.
- Category :
Answer
«Main Cause»
A lighting defect has occurred due to a broken filament in the D2 lamp or a fault in the constant current circuit, a lighting malfunction has occurred due to the lamp's life span, or there is a lighting defect due to a fault in the lamp itself.
«Countermeasure»
Replace the D2 lamp.
Click here for Replacement method of Lamp (HPLC)
If the error is not resolved, press the (power switch) to turn the power OFF and contact a Shimadzu sales or service representative.