Replacing the DL (LCMS)

Replacing the DL (LCMS)
WARNING: Before starting maintenance work, turn the heater OFF from the LabSolutions program and make sure that the temperature of the heated block has fallen to 50°C or lower.
The spray unit reaches high temperatures and could cause burns.
 
Please select a model.
Redoes from this question.
LCMS-8045(RX)/50(RX)/60(NX/RX)/65XE
Parts Number: LCMS-8060: 225-15718-43  DL ASSY3
Parts Number: LCMS-8050: 225-15718-42  DL ASSY2
Parts Number: LCMS-8045: 225-15718-91  DL ASSY
Redoes from this question.
Mounting the DL in the Instrument

Answer

1. Turn the notched section in the direction in which it engages (i.e., counterclockwise) to secure the DL to the IF flange(1), then connect the connector(2).
2. Tighten the two DL locking screws(3) with the hexagon wrench provided as an accessory.
    Tighten the locking screws sufficiently.
   
3. Insert the heater flange(4) into the IF flange(1) and tighten the two screws(5.
    Use the drawing jig to insert the heater flange.
   
    NOTE: When mounting the heater flange, take care about the positioning of the top and bottom of the flange.
4. Mount the ionization unit.
5. Lock the ionization unit by sliding its knob.
   