Maintenance of the Lens System (LCMS)

Maintenance of the Lens System (LCMS)
Since the lens system focuses the ions generated by the ionization unit, it becomes soiled over long periods of use and the detection sensitivity diminishes. For this reason you must clean the following lens system components: Qarray, skimmer, and multipoles. Also, clean the lens system components when a considerable decrease in sensitivity is observed.
 
Please select a model.
Redoes from this question.
LCMS-8045(RX)/8050(RX)/8060(NX/RX)/8065XE
Please select a maintenace item.
Redoes from this question.
Cleaning the Qarray

Answer

CAUTION: Do NOT disassemble the Qarray body. Disassembling them could cause injuries and instrument failure.
NOTE: Rest the Qarray on the stand and then remove the skimmer.
   
1. Clean the Qarray and skimmer.
  a) Remove the two screws(1) and remove the skimmer(2).
  b) Immerse the skimmer and Qarray in methanol and subject them to ultrasonic cleaning.
   
2. Return the parts to their original positions.
  a) Dry the parts after cleaning.
    NOTE: If the central part of the inner face of the skimmer is soiled, wipe it clean, e.g. with a cloth. If there is any contamination adhering to the central hole, remove it.
   
  b) Mount the skimmer(1) and secure it with the two screws(2) .
   