Cleaning the Interior of the ELSD
Cleaning the Interior of the ELSD
- Category :
Answer
1. Set the temperature to 100°C and flow gas at 350kPa.
2. Pump a suitable mobile phase at 2mL/min.
2. Pump a suitable mobile phase at 2mL/min.
CAUTION: Select a mobile phase in which the analyzed sample readily dissolves.
Avoid using any mobile phase that may damage the instrument.
If you cannot identify the sample, use ethanol or methanol.
Avoid using any mobile phase that may damage the instrument.
If you cannot identify the sample, use ethanol or methanol.
* With ELSD-LTII, after setting the drift tube temperature to 80 °C, you can set it to 100 °C by pressing and holding the "+" key.