• Font Size
  • S
  • M
  • L

Frequently Referenced FAQ

FAQ within [ Mobile Phase Monitor ]

1-6 of Total 6

≪ 1 / 1page ≫
  • Error Code:[0x5D01]

    «Main Cause» The remaining mobile phase volume is below the warning level. «Countermeasure» Replace the mobile phase. Show Detail

    • No:3840
    • Date of Issue:2020/12/02 16:43
  • Error Code:[0x5D02]

    «Main Cause» The remaining mobile phase volume is below the error level. «Countermeasure» Replace the mobile phase. Show Detail

    • No:3841
    • Date of Issue:2020/12/02 16:43
  • Error Code:[0x5D10]

    «Main Cause» A communication error occurred with the mobile phase monitor. «Countermeasure» Check that the bottle holder is properly connected and then unpl... Show Detail

    • No:3842
    • Date of Issue:2020/12/02 16:43
  • Error Code:[0x5D11]

    «Main Cause» The system was started without a previously connected bottle holder connected. Check that the sensors to be used are properly connected. «Count... Show Detail

    • No:3843
    • Date of Issue:2020/12/02 16:43
  • Error Code:[0x5D13]

    «Main Cause» The measurement value is out of range. «Countermeasure» Check that the bottle holder is properly connected and then unplug and reconnect the po... Show Detail

    • No:3844
    • Date of Issue:2020/12/02 16:43
  • Error Code:[0x5D14]

    «Main Cause» The zero point value has deviated significantly from the reference value. «Countermeasure» Please ensure there are no objects placed on the bot... Show Detail

    • No:3845
    • Date of Issue:2020/12/02 16:43

1-6 of Total 6

This site uses the FAQ system "PKSHA FAQ", which is a product of PKSHA Technology Inc.