• Font Size
  • S
  • M
  • L

Frequently Referenced FAQ

FAQ within [ System Controller ]

61-70 of Total 150

≪ 7 / 15page ≫
  • Error Code:[0x0082]

    «Main Cause» No Motor Home Position Error of RV.C has occurred. «Countermeasure» Turn the power OFF and then ON again. If the error persists, turn the powe... Show Detail

    • No:3729
    • Date of Issue:2020/12/02 16:31
  • Error Code:[0x0083]

    «Main Cause» No Motor Home Position Error of RV.D has occurred. «Countermeasure» Turn the power OFF and then ON again. If the error persists, turn the powe... Show Detail

    • No:3730
    • Date of Issue:2020/12/02 16:31
  • Error Code:[0x0084]

    «Main Cause» Valve home position error has occurred. «Countermeasure» Turn the power OFF and then ON again. If the error persists, turn the power OFF and c... Show Detail

    • No:3731
    • Date of Issue:2020/12/02 16:31
  • Error Code:[0x0085]

    «Main Cause» No Motor Home Position Error of Syringe unit has occurred. «Countermeasure» Turn the power OFF and then ON again. If the error persists, turn ... Show Detail

    • No:3732
    • Date of Issue:2020/12/02 16:31
  • Error Code:[0x0086]

    «Main Cause» The setting for suction/discharge of Syringe Unit is out of the range. «Countermeasure» Check the setting for suction/discharge of Syringe Unit. Show Detail

    • No:3733
    • Date of Issue:2020/12/02 16:31
  • Error Code:[0x008C]

    «Main Cause» SUBC: Leak sensor was detected leakage. «Countermeasure» Inspect the leakage from the flow line. Wipe out spilled liquid after fixing the leak... Show Detail

    • No:3734
    • Date of Issue:2020/12/02 16:31
  • Error Code:[0x000A]

    «Main Cause» Connection to a instrument with the [link check] checkbox selected wa disconnected. «Countermeasure» Please check the optical link cable. In pa... Show Detail

    • No:3767
    • Date of Issue:2020/12/02 16:43
  • Error Code:[0x2F01]

    «Main Cause» A communication error occurred with instrument. «Countermeasure» Check the optical link cable, turn off the main power of the system controller... Show Detail

    • No:3768
    • Date of Issue:2020/12/02 16:43
  • Error Code:[0x2F03]

    «Main Cause» A communication error occurred with instrument. «Countermeasure» Check the optical link cable, turn off the main power of the system controller... Show Detail

    • No:3769
    • Date of Issue:2020/12/02 16:43
  • Error Code:[0x2F05]

    «Main Cause» A communication error occurred with instrument. «Countermeasure» Check the optical link cable, turn off the main power of the system controller... Show Detail

    • No:3770
    • Date of Issue:2020/12/02 16:43

61-70 of Total 150

This site uses the FAQ system "PKSHA FAQ", which is a product of PKSHA Technology Inc.