FAQ - Frequently Asked Questions
- Products Search > HPLC > System Controller
Keyword Search
Frequently Referenced FAQ
FAQ within [ System Controller ]
-
«Main Cause» No Motor Home Position Error of RV.C has occurred. «Countermeasure» Turn the power OFF and then ON again. If the error persists, turn the powe... Show Detail
- No:3729
- Date of Issue:2020/12/02 16:31
-
«Main Cause» No Motor Home Position Error of RV.D has occurred. «Countermeasure» Turn the power OFF and then ON again. If the error persists, turn the powe... Show Detail
- No:3730
- Date of Issue:2020/12/02 16:31
-
«Main Cause» Valve home position error has occurred. «Countermeasure» Turn the power OFF and then ON again. If the error persists, turn the power OFF and c... Show Detail
- No:3731
- Date of Issue:2020/12/02 16:31
-
«Main Cause» No Motor Home Position Error of Syringe unit has occurred. «Countermeasure» Turn the power OFF and then ON again. If the error persists, turn ... Show Detail
- No:3732
- Date of Issue:2020/12/02 16:31
-
«Main Cause» The setting for suction/discharge of Syringe Unit is out of the range. «Countermeasure» Check the setting for suction/discharge of Syringe Unit. Show Detail
- No:3733
- Date of Issue:2020/12/02 16:31
-
«Main Cause» SUBC: Leak sensor was detected leakage. «Countermeasure» Inspect the leakage from the flow line. Wipe out spilled liquid after fixing the leak... Show Detail
- No:3734
- Date of Issue:2020/12/02 16:31
-
«Main Cause» Connection to a instrument with the [link check] checkbox selected wa disconnected. «Countermeasure» Please check the optical link cable. In pa... Show Detail
- No:3767
- Date of Issue:2020/12/02 16:43
-
«Main Cause» A communication error occurred with instrument. «Countermeasure» Check the optical link cable, turn off the main power of the system controller... Show Detail
- No:3768
- Date of Issue:2020/12/02 16:43
-
«Main Cause» A communication error occurred with instrument. «Countermeasure» Check the optical link cable, turn off the main power of the system controller... Show Detail
- No:3769
- Date of Issue:2020/12/02 16:43
-
«Main Cause» A communication error occurred with instrument. «Countermeasure» Check the optical link cable, turn off the main power of the system controller... Show Detail
- No:3770
- Date of Issue:2020/12/02 16:43
61-70 of Total 150