• Font Size
  • S
  • M
  • L

Frequently Referenced FAQ

FAQ within [ System Controller ]

51-60 of Total 150

≪ 6 / 15page ≫
  • Error Code:[0x1002]

    «Main Cause» Full-regeneration was operated while HPLC system was activated. «Countermeasure» Do not operate full-regeneration while HPLC system is activat... Show Detail

    • No:3550
    • Date of Issue:2020/12/02 16:30
  • Error Code:[0x1003]

    «Main Cause» The suppressor-related parameters are not correct. «Countermeasure» Check the values and set so that values are correct. Show Detail

    • No:3551
    • Date of Issue:2020/12/02 16:30
  • Error Code:[0x1004]

    «Main Cause» Suppressor cartridge current did not become constant. «Countermeasure» The suppressor cartridge may have become degraded. Replace the cartridge. Show Detail

    • No:3552
    • Date of Issue:2020/12/02 16:30
  • Error Code:[0x1005]

    «Main Cause» Suppressor cartridge current did not become constant. «Countermeasure» The suppressor cartridge may have become degraded. Replace the cartridge. Show Detail

    • No:3553
    • Date of Issue:2020/12/02 16:30
  • Error Code:[0x1006]

    «Main Cause» Suppressor cartridge current did not become constant. «Countermeasure» The suppressor cartridge may have become degraded. Replace the cartridge. Show Detail

    • No:3554
    • Date of Issue:2020/12/02 16:30
  • Error Code:[0x1007]

    «Main Cause» Suppressor cartridge current did not become constant. «Countermeasure» The suppressor cartridge may have become degraded. Replace the cartridge. Show Detail

    • No:3555
    • Date of Issue:2020/12/02 16:30
  • Error Code:[0x1009]

    «Main Cause» Column oven door was open for too long time while using suppressor. «Countermeasure» Close the door while using suppressor. Show Detail

    • No:3556
    • Date of Issue:2020/12/02 16:30
  • Error Code:[0x8632]

    «Main Cause» Cannot write to oven :Management Device (CMD). «Countermeasure» Once turn off the power, and turn on the power again. If this message still app... Show Detail

    • No:3627
    • Date of Issue:2020/12/02 16:30
  • Error Code:[0x0080]

    «Main Cause» No Motor Home Position Error of RV.A has occurred. «Countermeasure» Turn the power OFF and then ON again. If the error persists, turn the powe... Show Detail

    • No:3727
    • Date of Issue:2020/12/02 16:31
  • Error Code:[0x0081]

    «Main Cause» No Motor Home Position Error of RV.B has occurred. «Countermeasure» Turn the power OFF and then ON again. If the error persists, turn the powe... Show Detail

    • No:3728
    • Date of Issue:2020/12/02 16:31

51-60 of Total 150

This site uses the FAQ system "PKSHA FAQ", which is a product of PKSHA Technology Inc.