• Font Size
  • S
  • M
  • L

Frequently Referenced FAQ

FAQ within [ System Controller ]

121-130 of Total 150

13 / 15page
  • Error Code:[0x000B]

    «Main Cause» The system controller attempted to start an analysis without an instrument specified in [link check] settings being connected. «Countermeasure»... Show Detail

    • No:3821
    • Date of Issue:2020/12/02 16:43
  • Error Code:[0x000F]

    «Main Cause» A vial number specified in the sequence is out of range. «Countermeasure» Specify vial numbers that correspond to the given rack. Show Detail

    • No:3822
    • Date of Issue:2020/12/02 16:43
  • Error Code:[0x0019]

    «Main Cause» The autosampler was unable to recognize the rack or the rack is not located in the autosampler. «Countermeasure» Check the rack. Show Detail

    • No:3823
    • Date of Issue:2020/12/02 16:43
  • Error Code:[0x001D]

    «Main Cause» A vial number specified in the sequence is out of range. «Countermeasure» Specify vial numbers that correspond to the given rack. Show Detail

    • No:3824
    • Date of Issue:2020/12/02 16:43
  • Error Code:[0x4101]

    «Main Cause» (Only if used in the DHCP mode.) An IP address was not assigned by the DHCP server during startup. «Countermeasure» (1) Check the communication... Show Detail

    • No:3826
    • Date of Issue:2020/12/02 16:43
  • Error Code:[0x4102]

    «Main Cause» (Only if used in the DHCP mode.) An IP address was not assigned by the DHCP server when the IP address lease was extended. «Countermeasure» To ... Show Detail

    • No:3827
    • Date of Issue:2020/12/02 16:43
  • Error Code:[0x4103]

    «Main Cause» (Only if used in the DHCP mode.) An IP address was not assigned by the DHCP server when the IP address lease was extended. «Countermeasure» To ... Show Detail

    • No:3828
    • Date of Issue:2020/12/02 16:43
  • Error Code:[0x80F1]

    «Main Cause» The system controller cannot connect to the PDA detector due to a communication problem. «Countermeasure» (1) Check the communication environme... Show Detail

    • No:3829
    • Date of Issue:2020/12/02 16:43
  • Error Code:[0x80F2]

    «Main Cause» The system controller cannot connect to the PDA detector due to a communication problem. «Countermeasure» (1) Check the communication environme... Show Detail

    • No:3830
    • Date of Issue:2020/12/02 16:43
  • Error Code:[0x856A]

    «Main Cause» The injection volume exceeded the [injection volume upper limit] setting value. «Countermeasure» Inject a volume that does not exceed the [inje... Show Detail

    • No:3831
    • Date of Issue:2020/12/02 16:43

121-130 of Total 150

This site uses the FAQ system "PKSHA FAQ", which is a product of PKSHA Technology Inc.