Frequently Referenced FAQ

FAQ within [ System Controller ]

111-120 of Total 150

≪ 12 / 15page ≫
  • Error Code:[0x80F4]

    «Main Cause» The system name setting for the master PDA detector was changed during use. «Countermeasure» The parent-child relationship might have failed be... Show Detail

    • No:3811
    • Date of Issue:2020/12/02 16:43
  • Error Code:[0x80F5]

    «Main Cause» The PDA detector system name was changed during use. «Countermeasure» The parent-child relationship might have failed between the system contro... Show Detail

    • No:3812
    • Date of Issue:2020/12/02 16:43
  • Error Code:[0x80F7]

    «Main Cause» The PDA detector independently started up the system controller during an analysis. «Countermeasure» Either stop the PDA detector analysis or w... Show Detail

    • No:3813
    • Date of Issue:2020/12/02 16:43
  • Error Code:[0x8568]

    «Main Cause» The remaining volume of rinse solution is less than the check level. «Countermeasure» Replace the rinse solution or reset the rinse solution vo... Show Detail

    • No:3814
    • Date of Issue:2020/12/02 16:43
  • Error Code:[0x5C00]

    «Main Cause» The backup battery for the system controller is running low. «Countermeasure» Replace the battery early, because parameter settings cannot be s... Show Detail

    • No:3815
    • Date of Issue:2020/12/02 16:43
  • Error Code:[0x0003]

    «Main Cause» The method file, sequence file, and configuration settings were reinitialized (cleared). «Countermeasure» Check or reset method file, sequence ... Show Detail

    • No:3816
    • Date of Issue:2020/12/02 16:43
  • Error Code:[0x0004]

    «Main Cause» The method file, sequence file, and configuration settings were reinitialized (cleared). «Countermeasure» Check or reset method file, sequence ... Show Detail

    • No:3817
    • Date of Issue:2020/12/02 16:43
  • Error Code:[0x0005]

    «Main Cause» The method file, sequence file, and configuration settings were reinitialized (cleared). «Countermeasure» Check or reset method file, sequence ... Show Detail

    • No:3818
    • Date of Issue:2020/12/02 16:43
  • Error Code:[0x0006]

    «Main Cause» The method file, sequence file, and configuration settings were reinitialized (cleared). «Countermeasure» Check or reset method file, sequence ... Show Detail

    • No:3819
    • Date of Issue:2020/12/02 16:43
  • Error Code:[0x0008]

    «Main Cause» An error signal was input from an external device connected to the system controller external input/output terminal ([IN] terminal). «Counterme... Show Detail

    • No:3820
    • Date of Issue:2020/12/02 16:43